Non-invasive Instrument-Based Tests for Quantifying Anterior Chamber Flare in Uveitis: A Systematic Review
Authors:
Liu, X., McNally, T. W., Beese, S., Downie, L. E., Solebo, A. L., Faes, L., Husain, S., Keane, P. A., Moore, D. J., and Denniston, A. K.
Abstract:
Purpose: Anterior chamber (AC) flare is a key sign for anterior uveitis. New instrument-based techniques for measuring AC flare can offer automation and objectivity. This review aims to identify objective instrument-based measures for AC flare.Methods: In this systematic review, we identified studies reporting correlation between instrument-based tests versus clinician AC flare grading, and/or aqueous protein concentration, as well as test reliability.Results: Four index tests were identified in 11 studies: laser-flare photometry (LFP), optical coherence tomography, ocular flare analysis meter (OFAM) and the double-pass technique. The correlation between LFP and clinician grading was 0.40-0.93 and 0.87-0.94 for LFP and protein concentration. The double-pass technique showed no correlation with clinician grading and insufficient information was available for OFAM.Conclusion: LFP shows moderate to strong correlation with clinician grading and aqueous protein concentration. LFP could be a superior reference test compared to clinician AC flare grading for validating new index tests.